Fluke 725 Multifunction Process Calibrator |
The Fluke 725 Multifunction Process Calibrator is equipped to calibrate and test virtually any process parameter.
|
|
With the 725 you can source and measure temperature (RTDs and thermocouples), pressure using optional pressure modules, milliamps, ohms, frequency, and volts. The split display allows you to see output and input values at the same time when calibrating transmitters. For I/P and valve tests you can source milliamps while taking pressure measurements. The 725 has auto ramping and auto stepping for remote testing, plus 25 percent stepping for quick linearity tests.With its simple controls and memory functions, the 725 makes operation easy, sets up fast, and its durable design will give you many years of service.
725 Multifunction Process Calibrator Includes
725 Multifunction Process Calibrator Features
-
Compact design makes it easy to handle
-
Rugged design withstands rigors of the field
-
Easy to read source and measure screen lets you see output and input simultaneously
-
Measure milliamps, volts, RTDs, ohms, frequency, and thermocouples to test transmitters and sensors
-
Source and simulate milliamps, volts, RTDs, ohms, frequency, pressure, and thermocouples to calibrate transmitters
-
Measure and source pressure using any of 29 Fluke 700Pxx Pressure Modules
-
Source milliamps with simultaneous pressure measurement to conduct I/P and valve tests
-
Support flow meter testing with CPM and frequency functions
-
Perform quick linearity tests with auto ramp and auto step features
-
Power transmitters during test using loop supply with simultaneous milliamps measurement
-
Handling of fast pulsed RTD PLCs and transmitters, with pulses as short as 10 milliseconds
-
Store frequently used test setups for later recall
-
Backlight for poorly lit areas
-
Large capacity of four AA batteries
-
Battery door allows easy changes |
| Measurement Accuracy |
| Voltage DC |
| 30.000 V |
0.02%+ 2 counts (upper display) |
| 30.000 V |
0.02%+ 2 counts (lower display) |
| 100.00 mV |
0.02%+ 2 counts |
| -10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) | |
| Current DC |
| 24.000 mA |
0.02%+ 2 counts | |
| Resistance |
| 0.0 to 400.0 Ω |
0.1Ω (4-wire), 0.15Ω (2- and 3-wire) |
| 401 to 1500 Ω |
0.5Ω (4-wire), 1Ω (2- and 3-wire) |
| 1500 to 3200 Ω |
1Ω (4-wire), 1.5Ω (2- and 3-wire) | |
| Frequency |
| 2.0 to 1000.0 CPM |
0.05 % + 1 count |
| 1.0 to 1100.0 Hz |
0.05 % + 1 count |
| 1.00 to 10.00 kHz |
0.05 % + 1 count |
| Sensitivity |
1 V peak-to-peak-minimum | |
| Pressure |
| Accuracy |
from 0.025% of range using any of 29 pressure modules. |
| |
(for detailed specifications refer to pressure modules in options and accessories) |
| |
Modules available for differential, gage, vacuum, absolute, dual and high pressure. | |
| Source Accuracy |
| Voltage DC |
| 100.00 mV |
0.02%+2 counts |
| 10.000 V |
0.02%+2 counts |
| -10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) | |
| Current DC |
| 24.000 mA (Source) |
0.02%+ 2 counts |
| 24.000 mA (Simulate) |
0.02%+ 2 counts | |
| Resistance |
| 15.0 to 400.0 Ω |
0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
| 401 to 1500 Ω |
0.5 Ω(excitation current 0.05 to 0.8 mA) |
| 1500 to 3200 Ω |
1 Ω(excitation current 0.05 to 0.4 mA) | |
| Frequency |
| 2.0 to 1000.0 CPM |
0.05% |
| 1.0 to 1100.0 Hz |
0.05% |
| 1.00 to 10.00 kHz |
0.25% |
| Waveform: |
5 V p-p squarewave, -0.1 V offset | |
| RTDs and Thermocouples |
| Measure Accuracy: |
| NI-120 |
0.2 °C |
| PT-100 (385) |
0.33 °C |
| PT-100 (393) |
0.3 °C |
| PT-100 (JIS) |
0.3 °C |
| PT-200 (385) |
0.2 °C |
| PT-500 (385) |
0.3 °C |
| PT-1000 (385) |
0.2 °C |
| Resolution: |
0.1 °C |
| J: |
0.7 °C |
| K: |
0.8 °C |
| T: |
0.8 °C |
| E: |
0.7 °C |
| R: |
1.8 °C |
| S: |
1.5 °C |
| B: |
1.4 °C |
| L: |
0.7 °C |
| U: |
0.75 °C |
| N: |
0.9 °C |
| Resolution |
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
| XK |
0.6°C |
| BP |
1.2°C | |
| Source Accuracy: |
| NI-120 |
0.2 °C |
| PT-100 (385) |
0.33 °C |
| PT-100 (393) |
0.3 °C |
| PT-100 (JIS) |
0.3 °C |
| PT-200 (385) |
0.2 °C |
| PT-500 (385) |
0.3 °C |
| PT-1000 (385) |
0.2 °C |
| Resolution: |
0.1 °C |
| Note: |
Accuracy stated for 4-wire measurement. |
| J: |
0.7 °C |
| K: |
0.8 °C |
| T: |
0.8 °C |
| E: |
0.7 °C |
| R: |
1.4 °C |
| S: |
1.5 °C |
| B: |
1.4 °C |
| L: |
0.7 °C |
| U: |
0.75 °C |
| N: |
0.9 °C |
| Resolution |
J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C |
| XK |
0.6°C |
| BP |
1.2°C | |
| Specifications |
| Ramp Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Ramps |
Slow ramp, Fast ramp, 25% step-ramp | |
| Loop Power Function |
| Voltage: |
24 V |
| Accuracy: |
10% |
| Maximum current: |
22 mA, short circuit protected | |
| Step Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Steps |
25% of range, 100% of range | |
| Environmental Specifications |
| Operating Temperature |
|
| Storage Temperature |
|
| Operating Altitude |
|
| Safety Specifications |
| Agency Approvals |
| |
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | |
| Mechanical & General Specifications |
| Size |
|
| Weight |
|
| Batteries |
|
| Battery Replacement |
| |
Separate battery compartment, accessable without breaking calibration seal | |
| Side port connections |
| |
Pressure module connector, also used for remote real-time programming | | | |
|